Cepheid, Sample Collection Device, Part: 900-0370
Published August 21, 2024 · Recall initiated: June 20, 2024
Summary
FDA records medical device Class II recall Z-2584-2024, published August 21, 2024 and initiated June 20, 2024. Recalling firm: Cepheid; Status: Ongoing; Reported quantity: 651,150; Lot: UDI-DI: 28053326001523. Lot/Expiration: 230397900/ July 3, 2024; 230535300/ July 14, 2024; 230627500/ August 4, 2024; 231877400/ April 10, 2025. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Specimen collection device may leak after the patient sample swabs have been inserted into the test tubes, which could result in biohazard exposure, cross contamination causing false positive results for other specimens, and/or delay to test results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cepheid
- Distributor
- Worldwide distribution: US (nationwide): IL, TX, OH, LA, NY, CA, MD, IA, MN, KY, NH, WI, MO, NC, UT, WA, FL, NJ, SC, CO, PA, SD, NE, MT, WV, GA, MA, RI, CT, DE, MS, OK, VA, ID, AZ, OR, MI, IN, TN, AL, KS, ME, AR, NV, HI, AK, NM, WY, VT, ND, DC OUS (foreign) to countries of: Chile, Ireland, Oman, Nicaragua, Germany, San Marino, Ecuador, New Caledonia, France, Austria, United Kingdom, Finland, Italy, Hong Kong, Kuwait, United Arab Emirates, Netherlands, Belgium, Monaco, Colombia, Spain, Portugal, Poland, Switzerland, Denmark, Jersey, Mexico
- Lot
- UDI-DI: 28053326001523. Lot/Expiration: 230397900/ July 3, 2024; 230535300/ July 14, 2024; 230627500/ August 4, 2024; 231877400/ April 10, 2025
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Xpert® BCR-ABL Ultra Model/Catalog Number: GXBCRABL-10December 24, 2025
- FDAHeartMate 3 System Controllers provided within the following HeartMate 3 Left Ventricular Assist System (LVAS)August 21, 2024
- FDABD Kiestra Urine Culture Application Powered by BD Synapsys Informatics Solution; Catalog (Ref): 444900;August 21, 2024
- FDAVITROS Immunodiagnostic VITROS Folate Reagent Pack 1/2August 21, 2024
- FDABrand Name: UNclog Eye Mask Product Name: UNclog Eye Mask Model/Catalog Number: AG23JW24 Serial # ag22 ProductAugust 21, 2024
- FDAVarious Merit Medical convenience kits that contain recalled Jiangsu Shenli Medical Production Co. Ltd.August 21, 2024
Questions about this recall
Was Cepheid, Sample Collection Device, Part: 900-0370 recalled?
Yes. FDA published recall Z-2584-2024 on August 21, 2024. Reported hazard: Specimen collection device may leak after the patient sample swabs have been inserted into the test tubes, which could result in biohazard exposure, cross contamination causing false positive results for other specimens, and/or delay to test results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter