Back to search
FDAFDA · Food and drugs·Recall number Z-2587-2024

ClotTriever XL Catheter, 30 mm, REF: 41-102, used with Instructions for Use ClotTriever XL Catheter ClotTriever Thrombectomy System, Reference: IU-01011

Published August 28, 2024 · Recall initiated: July 19, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Thrombectomy system catheter instructions for use updated to reduce adverse events due to the following circumstances: 1) Fibrous, firmly adherent, calcified material, or suspected tumor thrombus overloading catheter bag 2) Catheter pulled through internal jugular vein access site or not pulled away from heart; may lead to device entrapment, pulmonary embolism, venous cutdown, abdominal surgery.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Inari Medical - Oak Canyon
Distributor
Worldwide - US Nationwide distribution including in the states of PA, FL, GA, CA, IL, SC, NY, MN, UT, TN, WI, MI, KS, SD, AZ, AR, IN, KY, TX, AL, MS, DE, MA, OH, MO, WV, VA, NC, NV, CO, NE, OR, NM, NH, NJ, CT, ID, IA, WA, MD, OK, RI, DC, PR, LA, ND, VT and the countries of New Zealand, Chile, Australia.
Lot
UDI-DI:00850291007277. All lots. IU-01011 REV. C.
Risk classification
Class I
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card