Adult ThermoGard® Dual Dispersive Electrodes (for patients >15 kg.), 10' (3.05m) Cable, Catalog Number 51-7310. Intended to be used for the dispersion and return to the electrosurgical generator.
Published August 31, 2016 · Recall initiated: May 18, 2016
Summary
FDA records medical device Class II recall Z-2595-2016, published August 31, 2016 and initiated May 18, 2016. Recalling firm: Conmed Corporation; Status: Terminated; Reported quantity: 307,684 units; Lot: Beginning Lot Code 141001X with Ending Lot Code 20150705X. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Conmed Corporation
- Distributor
- Worldwide Distribution -- USA, including AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, & Puerto Rico; and, the countries of Australia, Canada, Belgium, Brunei Darussalam, China, France, Guatemala, Hong Kong, Israel, Italy, Japan, South Korea, Malaysia, Mexico, New Zealand, Norway, Portugal, Saudi Arabia, Spain, Sweden, and United Arab Emirates.
- Lot
- Beginning Lot Code 141001X with Ending Lot Code 20150705X.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAConMed Anchor Tissue Retrieval System 15 MM, 1550 ML (3/BX) Catalog Number: TRS175SB2 The Anchor TissueMay 15, 2019
- FDAConMed Anchor Tissue Retrieval System 10 MM, 235 ML, (5/BX) Catalog Number: TRS100SB2 The Anchor TissueMay 15, 2019
- FDAAlere INRatio PT/INR Test Strips Model Number: 0100071, 0100139 Product Usage: The INRatio/INRatio2 TestAugust 31, 2016
- FDALarge QWIX Compression screw, dia. 7.5 mm L 90 mm, Lot No. FH86, Rx only, Product Usage: The QWIX PositioningAugust 31, 2016
- FDAAlere INRatio2 PT/INR Professional Monitoring System Model Number: 55128A Product Usage: The AlereAugust 31, 2016
- FDARadiesse (+) Lidocaine, Injectable Implant 1.5cc, Rx only, P/N: 8063M0K1, Merz Aesthetics. Product Usage:August 31, 2016
Questions about this recall
Was Adult ThermoGard® Dual Dispersive Electrodes (for patients >15 kg.), 10' (3.05m) Cable, Catalog Number 51-7310. Intended to be used for the dispersion and return to the electrosurgical generator. recalled?
Yes. FDA published recall Z-2595-2016 on August 31, 2016. Reported hazard: A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.
What should I do if I have this product?
Voluntary: Firm initiated · Letter