Adult ThermoGard® Dual Dispersive Electrodes (for patients >15 kg.), 10' (3.05m) Cable, Catalog Number 51-7310. Intended to be used for the dispersion and return to the electrosurgical generator.

Published August 31, 2016 · Recall initiated: May 18, 2016

Summary

FDA records medical device Class II recall Z-2595-2016, published August 31, 2016 and initiated May 18, 2016. Recalling firm: Conmed Corporation; Status: Terminated; Reported quantity: 307,684 units; Lot: Beginning Lot Code 141001X with Ending Lot Code 20150705X. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Conmed Corporation
Distributor
Worldwide Distribution -- USA, including AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, & Puerto Rico; and, the countries of Australia, Canada, Belgium, Brunei Darussalam, China, France, Guatemala, Hong Kong, Israel, Italy, Japan, South Korea, Malaysia, Mexico, New Zealand, Norway, Portugal, Saudi Arabia, Spain, Sweden, and United Arab Emirates.
Lot
Beginning Lot Code 141001X with Ending Lot Code 20150705X.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Adult ThermoGard® Dual Dispersive Electrodes (for patients >15 kg.), 10' (3.05m) Cable, Catalog Number 51-7310. Intended to be used for the dispersion and return to the electrosurgical generator. recalled?

Yes. FDA published recall Z-2595-2016 on August 31, 2016. Reported hazard: A version of the dual dispersive electrodes may not be compatible with some electrosurgical generators, placing patients at risk for undetected pad lift and a potential burn.

What should I do if I have this product?

Voluntary: Firm initiated · Letter