Dimension Vista Thyroid Stimulating Hormone (TSH) Flex reagent cartridge

Published June 28, 2017 · Recall initiated: March 21, 2017

Summary

FDA records medical device Class II recall Z-2611-2017, published June 28, 2017 and initiated March 21, 2017. Recalling firm: Siemens Healthcare Diagnostics, Inc.; Status: Terminated; Reported quantity: 19, 927.00; Lot: Device Listing Number # D230089 Lot # 16118BD, 16159BA,16193BC,.16193BD,16208AA,16251BA 16277BD,16322BB,17005BB,17023AC. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Healthcare Diagnostics, Inc.
Distributor
Product was distributed throughout the United States and Canada, United Arab Emirates, Argentina, Austria,Australia,Bangladesh,Belgium,Brazil,Canada Switzerland,China,Columbia,Czech Republic,Germany,Denmark,Algeria Egypt,Spain,Finland,France,Great Britain,Greece,Croatia,Indonesia,Ireland India,Italy,Japan,Canary Islands,Republic of Korea,Kuwait,Lebanon,Latvia Macedonia,Netherlands,Norway,New Zealand,Philippines,Poland,Portugal Qatar,Saudi Arabia,Slovenia,Slovakia,Turkey,Taiwan,South Africa
Lot
Device Listing Number # D230089 Lot # 16118BD, 16159BA,16193BC,.16193BD,16208AA,16251BA 16277BD,16322BB,17005BB,17023AC
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Dimension Vista Thyroid Stimulating Hormone (TSH) Flex reagent cartridge recalled?

Yes. FDA published recall Z-2611-2017 on June 28, 2017. Reported hazard: The concentrations for Biotin listed in the Non-Interfering Substances section of the current Dimension and Dimension Vistat Instructions for Use (IFU) for Dimension TNI, Dimension Vista DIGXN, E2, FERR, PRL, and TSH have incorrect units, and incorrectly state the level at which biotin does not interfere.

What should I do if I have this product?

Voluntary: Firm initiated · Letter