Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator
Published July 8, 2026 · Recall initiated: June 16, 2026
Summary
FDA records medical device Class II recall Z-2644-2026, published July 8, 2026 and initiated June 16, 2026. Recalling firm: Accuray Incorporated; Status: Ongoing; Reported quantity: 292 units; Lot: ALL Cyberknife systems running software version 11.X and higher that utilize the Exchange Table. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Accuray Incorporated
- Distributor
- Worldwide - US Nationwide and and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kenya, Korea, Republic of, Liechtenstein, Lithuania, Luxembourg, Macao, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
- Lot
- ALL Cyberknife systems running software version 11.X and higher that utilize the Exchange Table.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear AcceleratorJuly 8, 2026
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDABrand Name: Medela Product Name: ENFit ExSet Strl 60 2.0ml PV 1ct EN Model/Catalog Number: ENF060182LD /July 8, 2026
- FDACyberKnife Treatment Delivery System - Product Usage: indicated for image-guided stereotactic radiosurgery andMarch 24, 2021
- FDACyberKnife Robotic Radiosurgery Systems, models: G3, G4, VSI. Radiology: The CyberKnife Robotic RadiosurgeryDecember 28, 2016
- FDACyberKnife Robotic Radiosurgery System with the first generation IRIS Variable Aperture Collimator. TheNovember 26, 2014
Questions about this recall
Was Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator recalled?
Yes. FDA published recall Z-2644-2026 on July 8, 2026. Reported hazard: In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.
What should I do if I have this product?
Voluntary: Firm initiated · Letter