Back to search
FDAFDA · Food and drugs·Recall number Z-2644-2026

Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator

Published July 8, 2026 · Recall initiated: June 16, 2026

Summary

FDA records medical device Class II recall Z-2644-2026, published July 8, 2026 and initiated June 16, 2026. Recalling firm: Accuray Incorporated; Status: Ongoing; Reported quantity: 292 units; Lot: ALL Cyberknife systems running software version 11.X and higher that utilize the Exchange Table. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Accuray Incorporated
Distributor
Worldwide - US Nationwide and and the countries of Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, France, Germany, Greece, Hong Kong, India, Ireland, Italy, Japan, Jordan, Kenya, Korea, Republic of, Liechtenstein, Lithuania, Luxembourg, Macao, Mexico, Netherlands, New Zealand, Pakistan, Poland, Portugal, Qatar, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom.
Lot
ALL Cyberknife systems running software version 11.X and higher that utilize the Exchange Table.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Cyberknife Treatment Deliver System equipped with the Xchange Robotic Collimator Changer; Linear Accelerator recalled?

Yes. FDA published recall Z-2644-2026 on July 8, 2026. Reported hazard: In certain cases, a collimator may not fully dock, which triggers a system interlock and prevents motion from the operator console. If personnel then enter the treatment room and initiate motion using the in-room control pendant, the collimator may release and fall from the housing.

What should I do if I have this product?

Voluntary: Firm initiated · Letter