SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G (UDI 00801741087639), PA-0033G (UDI 00801741096433)
Published July 29, 2020 · Recall initiated: June 19, 2020
Summary
FDA records medical device Class II recall Z-2668-2020, published July 29, 2020 and initiated June 19, 2020. Recalling firm: Becton Dickinson & Company; Status: Terminated; Reported quantity: 5,296,132 total; Lot: Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G (UDI 00801741087639), PA-0033G (UDI 00801741096433). Lot Numbers: REEN3159 REEP3067 REEQ4064 REDS3541 REDV3438 REEN2781 REEN2145 REEP3068 REDY0835 REDZ3103. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Becton Dickinson & Company
- Distributor
- Worldwide Distribution
- Lot
- Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G (UDI 00801741087639), PA-0033G (UDI 00801741096433). Lot Numbers: REEN3159 REEP3067 REEQ4064 REDS3541 REDV3438 REEN2781 REEN2145 REEP3068 REDY0835 REDZ3103
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Quincke Needle 26 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 22 G x 3.5 in.; lidocaine HCL (1%), 5 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDABD®Spinal Tray with BD®Whitacre Needle 25 G x 3.5 in.; lidocaine HCL (1%), 2 mL, bupivacaine HCL (0.75%) withJune 10, 2026
- FDASafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing; Catalog Number CPA00040 (UDIJuly 29, 2020
- FDASafeStep Huber Needle Set with Y Injection Site and Medegen Needleless Injection Cap; Catalog NumbersJuly 29, 2020
Questions about this recall
Was SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G (UDI 00801741087639), PA-0033G (UDI 00801741096433) recalled?
Yes. FDA published recall Z-2668-2020 on July 29, 2020. Reported hazard: Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.
What should I do if I have this product?
Voluntary: Firm initiated · Letter