SafeStep Huber Needle Set with GuardIVa Antimicrobial Hemostatic Dressing Port Access Kit; Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G (UDI 00801741087639), PA-0033G (UDI 00801741096433)

Published July 29, 2020 · Recall initiated: June 19, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential development of cracks or breaks in the tubing near the Luer or Y site of the device. Devices that have developed breaks or cracks can cause leakage resulting in exposure to chemicals such a chemotherapeutics or biohazard material (e.g., blood). A crack or break could result in potential blood loss, catheter occlusion, air embolism, under or interrupted infusion, or site contamination which could lead to infection.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Becton Dickinson & Company
Distributor
Worldwide Distribution
Lot
Catalog Numbers PA-0031G (UDI 00801741085666), PA-0031YNG (UDI 00801741085673), PA-0032G (UDI 00801741087639), PA-0033G (UDI 00801741096433). Lot Numbers: REEN3159 REEP3067 REEQ4064 REDS3541 REDV3438 REEN2781 REEN2145 REEP3068 REDY0835 REDZ3103
Risk classification
Class II
Area (FDA)
device
Country
United States