DePuy Synthes TI Vectra Plates: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate 3 Level / 57 mm part 04.613.257 lot 3693878 TI Vectra" Plate 4 Level / 64 mm part 04.613.364 lot 8143905
Published October 1, 2014 · Recall initiated: September 2, 2014
Summary
FDA records medical device Class II recall Z-2674-2014, published October 1, 2014 and initiated September 2, 2014. Recalling firm: Synthes, Inc.; Status: Terminated; Reported quantity: 16; Lot: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate 3 Level / 57 mm part 04.613.257 lot 3693878 TI Vectra" Plate 4 Level / 64 mm part 04.613.364 lot 8143905. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
DePuy Synthes is initiating a voluntary medical device recall of certain lots of the TI Vectra Plates, which is a part of the DePuy Synthes Vectra Anterior Cervical Plate Family of Systems. The Vectra Systems are intended for anterior screw fixation to the cervical spine (C2-C7). It was discovered in certain lots, that the clips are potentially missing from the TI Vectra Plates resulting in the i
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes, Inc.
- Distributor
- US Distribution including the states of IL, SC, VA, TX, MA, NC, LA, CA, AL, CO, NE, OR and KS., and Internationally to Canada.
- Lot
- TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate 3 Level / 57 mm part 04.613.257 lot 3693878 TI Vectra" Plate 4 Level / 64 mm part 04.613.364 lot 8143905
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 10MM STERILE, Part Number 04.210.110TS - Product Usage:Sterile TubeFebruary 5, 2020
- FDA2.4MM TI VA LOCKING SCREW STARDRIVE 18MM STERILE, Part Number 04.210.118TS - Product Usage:Sterile TubeFebruary 5, 2020
- FDAPull Reduction Device For 4.3mm Percutaneous Drill Guide; The pull reduction device with quick coupling isJune 24, 2015
- FDA3.7mm Cannulated Locking Screws and 3.7mm Cannulated Conical Screws; . the Cannulated Screws are indicated forJune 24, 2015
- FDABD PosiFlush SF Saline Flush Syringe 0.9 Sodium Chloride Injection, USP 10 mL REF 306553. Intended to be usedOctober 1, 2014
- FDAThe synthetic non-absorbable surgical sutures CARDIOXYL are composed of a polyester [poly(ethyleneOctober 1, 2014
Questions about this recall
Was DePuy Synthes TI Vectra Plates: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" Plate 3 Level / 57 mm part 04.613.257 lot 3693878 TI Vectra" Plate 4 Level / 64 mm part 04.613.364 lot 8143905 recalled?
Yes. FDA published recall Z-2674-2014 on October 1, 2014. Reported hazard: DePuy Synthes is initiating a voluntary medical device recall of certain lots of the TI Vectra Plates, which is a part of the DePuy Synthes Vectra Anterior Cervical Plate Family of Systems. The Vectra Systems are intended for anterior screw fixation to the cervical spine (C2-C7). It was discovered in certain lots, that the clips are potentially missing from the TI Vectra Plates resulting in the i
What should I do if I have this product?
Voluntary: Firm initiated · Letter