Circular Lock Screws, part # G826, lot # 11406. The Lock Screws are part of the Genesys Spine TiLock Pedicle Screw System. The TiLock Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion).
Published October 1, 2014 · Recall initiated: April 2, 2014
Summary
FDA records medical device Class II recall Z-2677-2014, published October 1, 2014 and initiated April 2, 2014. Recalling firm: Genesys Orthopedics Systems, LLC; Status: Terminated; Reported quantity: 642; Lot: The devices being recalled have a catalog number of G826 and a lot number of 11406. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The recall is being initiated because MK Precision (the Circular Lock Screw supplier) notified Genesys Spine that a portion (less than 20%) of the Circular Lock Screws in lot 11406 were manufactured using Stainless Steel in the center Post instead of the specified titanium alloy.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Genesys Orthopedics Systems, LLC
- Distributor
- Nationwide Distribution including NV, CA, NY, MO, TX, FL, NJ, GA, TN.
- Lot
- The devices being recalled have a catalog number of G826 and a lot number of 11406.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe synthetic non-absorbable surgical sutures CARDIOFLON Evolution are composed of polyester [poly(ethyleneOctober 1, 2014
- FDANo packaging. The device is a computed tomography scanner. Intended to provide corss sectional images of theOctober 1, 2014
- FDAThe synthetic non-absorbable surgical sutures CARDIOXYL are composed of a polyester [poly(ethyleneOctober 1, 2014
- FDAMini Cannulated Titanium Headed and Headless Screw Set and 2.5, 3.0, and 4.0 mm countersink/depth gauges inOctober 1, 2014
- FDADePuy Synthes TI Vectra Plates: TI Vectra-One" Plate 2 Level / 34 mm part 04.613.184 3337672 TI Vectra" PlateOctober 1, 2014
- FDAThe Ysio Systems with software version VC10 The Ysio enables radiographic and tomographic exposures of theOctober 1, 2014
Questions about this recall
Was Circular Lock Screws, part # G826, lot # 11406. The Lock Screws are part of the Genesys Spine TiLock Pedicle Screw System. The TiLock Pedicle Screw System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of thoracic, lumbar, and sacral spine: degenerative disc disease (DDD) (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e. fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion). recalled?
Yes. FDA published recall Z-2677-2014 on October 1, 2014. Reported hazard: The recall is being initiated because MK Precision (the Circular Lock Screw supplier) notified Genesys Spine that a portion (less than 20%) of the Circular Lock Screws in lot 11406 were manufactured using Stainless Steel in the center Post instead of the specified titanium alloy.
What should I do if I have this product?
Voluntary: Firm initiated · Letter