Leksell Gamma Knife Icon is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g., metastatic tumors, recurrent glioblastomas, trigeminal neuralgia, medically refractory essential tremor, orbital tumors, ocular tumors, optic nerve tumors, benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, cavernous malformations, chordomas, glomus tumors, hemangiomas), skull base tumors, head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, oropharynx, nasopharynx, sinonasal, salivary gland), and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, cavernous malformations, skull base tumors).
Published July 19, 2017 · Recall initiated: May 23, 2017
Summary
FDA records medical device Class II recall Z-2697-2017, published July 19, 2017 and initiated May 23, 2017. Recalling firm: Elekta, Inc.; Status: Terminated; Reported quantity: 2; Lot: Serial Numbers 6018, 6047, 6090, 6094, 6171, 8018. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If the gantry module is replaced after original installation the configuration settings may be missing. The identified risk for this issue is electrical safety for technicians doing maintenance on the X-ray generator.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Elekta, Inc.
- Distributor
- US Distribution to the states of : NY and PA.
- Lot
- Serial Numbers 6018, 6047, 6090, 6094, 6171, 8018
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABrand Name: Geneva Product Name: Brachytherapy Applicator Product Description: The device is intended forJuly 15, 2026
- FDABrand Name: Fletcher CT/MR Shielded Applicator Set Product Name: Intracavitary Remote Afterloading ApplicatorJuly 15, 2026
- FDABrand Name: Brachytherapy Applicator Product Name: Advanced Gynecological Applicator Product Description: TheJuly 15, 2026
- FDAMOSAIQ Oncology Information SystemJune 17, 2026
- FDAInfinity 7, Implantable Pulse Generator, REF 6663, Rx only, STERILE EO Product Usage: The Infinity IPG is anJuly 19, 2017
- FDAArrow EVOLUTION Pressure Injectable PICC with Chlorag+ard Antimicrobial and Antithrombogenic TechnologyJuly 19, 2017
Questions about this recall
Was Leksell Gamma Knife Icon is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g., metastatic tumors, recurrent glioblastomas, trigeminal neuralgia, medically refractory essential tremor, orbital tumors, ocular tumors, optic nerve tumors, benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, cavernous malformations, chordomas, glomus tumors, hemangiomas), skull base tumors, head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, oropharynx, nasopharynx, sinonasal, salivary gland), and pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, cavernous malformations, skull base tumors). recalled?
Yes. FDA published recall Z-2697-2017 on July 19, 2017. Reported hazard: If the gantry module is replaced after original installation the configuration settings may be missing. The identified risk for this issue is electrical safety for technicians doing maintenance on the X-ray generator.
What should I do if I have this product?
Voluntary: Firm initiated · Letter