MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896

Published July 15, 2026 · Recall initiated: May 27, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Medtronic MiniMed, Inc.
Distributor
International distribution to the countries of Spain, Great Britain, Netherlands, Italy.
Lot
UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62
Risk classification
Class II
Area (FDA)
device
Country
United States