MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
Published July 15, 2026 · Recall initiated: May 27, 2026
Summary
FDA records medical device Class II recall Z-2698-2026, published July 15, 2026 and initiated May 27, 2026. Recalling firm: Medtronic MiniMed, Inc.; Status: Ongoing; Reported quantity: 778; Lot: UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
What to do
Voluntary: Firm initiated
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic MiniMed, Inc.
- Distributor
- International distribution to the countries of Spain, Great Britain, Netherlands, Italy.
- Lot
- UDI-DI/GTIN: A7630006392701, 00763000521554, 00763000639273, A7630006392301, 00199150035687, A7630007325603, A7630007325601, 00763000639235, A7630007325604, 00763000384289, 00763000521530, A7630007325401, 00763000521547, A7630006392601, A7630006392201, 00763000639228, 00199150035663, 00763000639266, A7630007325404, 00199150035694, 00763000442194, A7630007325403, 00763000521523. Software version 6.62
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGeneric Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601October 28, 2026
- FDACareLink Clinic, REF: MMT-7350October 28, 2026
- FDABrand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072October 14, 2026
- FDAMiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLWAugust 26, 2026
- FDAMedline Convenience Kits: 1) OPEN HEART PEDS PACK, Model Number: DYNJ88569; 2) OPEN HEART PEDS PACK, ModelJuly 15, 2026
- FDAKL Owen Custom Orthodontic Solution, model/catalog number CIAS-800033,July 15, 2026
Questions about this recall
Was MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896 recalled?
Yes. FDA published recall Z-2698-2026 on July 15, 2026. Reported hazard: A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
What should I do if I have this product?
Voluntary: Firm initiated