Optima XR240amx, Mobile Digital Radiographic System

Published July 22, 2026 · Recall initiated: June 5, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
GE Medical Systems, LLC
Distributor
Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.
Lot
UDI-DI: 00840682122351 and 00840682123686; System IDs: 252847PP249, 281MSL2402, 291OPTI1, 312CCHSUPER7, 318487OPTIMA1, 360744OPT2402, 501686OPTI1, 504826XR241, 519941XR240A, 540776ODP3, 541472M4, 602344240C, 706291BEE9, 713799XR240, 770836BEE2, 813872OPT11, 813872OPT9, 843724OPT1, 843847OPT240, 870239OPT2, 905522AMX240, 973625DIG4, DF2402001002WK, SA1013MM13
Risk classification
Class II
Area (FDA)
device
Country
United States