FDAFDA · Food and drugs·Recall number Z-2741-2020
Alaris System PC Unit Model 8015. modular infusion pump and monitoring system
Published August 12, 2020 · Recall initiated: June 30, 2020
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- CareFusion 303, Inc.
- Distributor
- U.S.: CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, KS, MN, MI, WA, UT, MD, TN, CO, NJ, NE, ND, WV, KY, IA, MS, AL, DE, AK, VA, OK, NM, GA, CT, SD, NV, ID, MT, ME, VT, WY, NH, MP, DC, GU, PR, AS, RI. O.U.S.: TR, QA, CA, AE, KW, MX, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, BH
- Lot
- All serial numbers
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States