Alaris System PC Unit Model 8015. modular infusion pump and monitoring system

Published August 12, 2020 · Recall initiated: June 30, 2020

Summary

FDA records medical device Class I recall Z-2741-2020, published August 12, 2020 and initiated June 30, 2020. Recalling firm: CareFusion 303, Inc.; Status: Terminated; Reported quantity: 824277; Lot: All serial numbers. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
CareFusion 303, Inc.
Distributor
U.S.: CO, FL, SC, NY, CA, NC, TX, IL, PA, AZ, LA, IN, MA, MO, AR, WI, HI, OR, OH, KS, MN, MI, WA, UT, MD, TN, CO, NJ, NE, ND, WV, KY, IA, MS, AL, DE, AK, VA, OK, NM, GA, CT, SD, NV, ID, MT, ME, VT, WY, NH, MP, DC, GU, PR, AS, RI. O.U.S.: TR, QA, CA, AE, KW, MX, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, BH
Lot
All serial numbers
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Alaris System PC Unit Model 8015. modular infusion pump and monitoring system recalled?

Yes. FDA published recall Z-2741-2020 on August 12, 2020. Reported hazard: If one or more screws or washers are loose or missing causing the battery not to be properly secured to the PC Unit, that is running on battery power, the system may experience a power loss with a prolonged, non-silenceable alarm. Power loss may result in an interruption of patient therapy or monitoring.