FDAFDA · Food and drugs·Recall number Z-2742-2015

CDI 500 Blood Parameter Monitoring System. Provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature.

Published September 16, 2015 · Recall initiated: August 7, 2015

Summary

FDA records medical device Class II recall Z-2742-2015, published September 16, 2015 and initiated August 7, 2015. Recalling firm: Terumo Cardiovascular Systems Corporation; Status: Terminated; Reported quantity: 4638; Lot: 500A, 500AHCT, 500AV, 500AVHCT CDI Blood Parameter Monitoring System 500; software version 1.69. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Inaccuracies in SvO2, temperature, pH, pCO2, pO2, Hematocrit, and Potassium readings following a software upgrade to version 1.69.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Terumo Cardiovascular Systems Corporation
Distributor
Worldwide Distribution-US (nationwide) including DC and PR and the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and the countries of Australia, Belgium, Canada, Chile, Colombia, Hong Kong, Indonesia, Japan, Malaysia, Mexico, Philippines, Singapore, South Korea, Taiwan, Thailand, United Arab Emirates.
Lot
500A, 500AHCT, 500AV, 500AVHCT CDI Blood Parameter Monitoring System 500; software version 1.69.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was CDI 500 Blood Parameter Monitoring System. Provides continuous, on-line monitoring of the extracorporeal partial pressure of oxygen and and carbon dioxide, pH, potassium, oxygen saturation, hematocrit, hemoglobin and temperature. recalled?

Yes. FDA published recall Z-2742-2015 on September 16, 2015. Reported hazard: Inaccuracies in SvO2, temperature, pH, pCO2, pO2, Hematocrit, and Potassium readings following a software upgrade to version 1.69.

What should I do if I have this product?

Voluntary: Firm initiated · Letter