Atrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85355 (Marketed OUTSIDE U.S. only) The iCAST Covered Stent is indicated for the treatment of tracheobronchial strictures produced by malignant neoplasms.
Published August 2, 2017 · Recall initiated: May 25, 2017
Summary
FDA records medical device Class II recall Z-2761-2017, published August 2, 2017 and initiated May 25, 2017. Recalling firm: Atrium Medical Corporation; Status: Terminated; Reported quantity: 246 (US: 183 units; Foreign: 63 units); Lot: Part Number 85455: Lots 240735, 240523; Part Number 85355: Lot 240734. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm received a complaint that a customer who ordered an ADVANTA V12 7mm x 22mm x 120cm received an ADVANTA v12 7 mm x 38 mm x 80 cm, due to the manifold (catheter hub) was printed as 7mm x 38 mm x 80 cm. The complaint was the result of incorrect dimensional information (7mm x 38mm x 80cm) printed on the manifiold (catheter hub). The complaint device physical dimensions matched the dimensional information displayed on the outer packaging and the device inner pouch. This was the only complaint received for this non-conformance.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Atrium Medical Corporation
- Distributor
- Worldwide distribution-US (nationwide) and countries of: Germany, Great Britain, Austria, Ireland, Netherlands, France, Italy, Sweden, and San Marino.
- Lot
- Part Number 85455: Lots 240735, 240523; Part Number 85355: Lot 240734
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAGetinge PVC Thoracic Catheter, Soft PVC Right Angle, 10 per Case; Size(Model Number): 1. 24FR(8124), 2.September 16, 2026
- FDAGetinge PVC Thoracic Catheter, Firm PVC Straight, 10 per Case; Size(Model Number): 1. 28FR(15028), 2.September 16, 2026
- FDAGetinge PVC Thoracic Catheter, Soft PVC Straight, 10 per Case; Size(Model Number): 1. 16FR(8016), 2.September 16, 2026
- FDAiCast Covered Stent, 6MMx16MMx120CM, Model Number 42616October 16, 2024
- FDASmiths Medical CADD Blue-Striped IV Administration Set, REF 21-7036-01, Sterilized using ethylene oxide.August 2, 2017
- FDAPacific Hemostasis FDP (Fibrinogen Degradation Products) Assay Kit (30 Determinations), REF/Model 100650, UDIAugust 2, 2017
Questions about this recall
Was Atrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85355 (Marketed OUTSIDE U.S. only) The iCAST Covered Stent is indicated for the treatment of tracheobronchial strictures produced by malignant neoplasms. recalled?
Yes. FDA published recall Z-2761-2017 on August 2, 2017. Reported hazard: Firm received a complaint that a customer who ordered an ADVANTA V12 7mm x 22mm x 120cm received an ADVANTA v12 7 mm x 38 mm x 80 cm, due to the manifold (catheter hub) was printed as 7mm x 38 mm x 80 cm. The complaint was the result of incorrect dimensional information (7mm x 38mm x 80cm) printed on the manifiold (catheter hub). The complaint device physical dimensions matched the dimensional information displayed on the outer packaging and the device inner pouch. This was the only complaint received for this non-conformance.
What should I do if I have this product?
Voluntary: Firm initiated · Letter