Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application.
Published July 29, 2026 · Recall initiated: May 29, 2026
Summary
FDA records medical device Class II recall Z-2777-2026, published July 29, 2026 and initiated May 29, 2026. Recalling firm: Emano Metrics, Inc.; Status: Ongoing; Reported quantity: 685; Lot: Software Version - 2.0.0 GS1 GTIN code: 199874903484 UDI code: GUDID registration is in progress. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Emano Metrics, Inc.
- Distributor
- U.S. Nationwide distribution in the states of AR, AZ, CA, CO, FL. IA, ID, NC, NJ, PA, TX and WI.
- UPC
- 199874903484
- Lot
- Software Version - 2.0.0 GS1 GTIN code: 199874903484 UDI code: GUDID registration is in progress
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOlympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4July 29, 2026
- FDAMiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)July 29, 2026
- FDASnapshotGLO Wound Imaging Device, Model KB100, KB100AA.July 29, 2026
- FDAMedline convenience kits containing Bard® Silastic® Foley Catheters (French sizes 16, 18, 20) 5cc balloons.July 29, 2026
- FDAPipeline Flex Embolization Device with Shield Technology, 3.00 mm x 20 mm 027 REF PED2-300-20. IntracranialJuly 29, 2026
- FDAMedline convenience kits containing Bard® Silastic® Foley Catheters (French sizes 16, 18, 20) 5cc balloons.July 29, 2026
Questions about this recall
Was Emano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acoustic uroflowmetry system. Software-only device that measures urinary flow rate and voided volume from the acoustic signal of urination. No physical components. Distributed digitally via mobile application. recalled?
Yes. FDA published recall Z-2777-2026 on July 29, 2026. Reported hazard: Problem description/Reason for recall: During the period March 22, 2026, to May 29, 2026, recorded urination audio was captured at 48,000 Hz while the model was trained and validated on 44,100 Hz audio. The models depend on the sampling so every recording in the window was therefore processed against a time/frequency scale the models did not expect, producing systematically biased Qmax, Qmean, and voided-volume estimates. The problem usually over-reports flow rate and volume; the firm have made corrected results available to care teams.
What should I do if I have this product?
Voluntary: Firm initiated · Letter
How do I know if my unit is part of the recall?
Compare your product details with the notice. UPC: 199874903484.