MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)
Published July 29, 2026 · Recall initiated: June 16, 2026
Summary
FDA records medical device Class II recall Z-2778-2026, published July 29, 2026 and initiated June 16, 2026. Recalling firm: Medtronic MiniMed, Inc.; Status: Ongoing; Reported quantity: 3,465; Lot: Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic MiniMed, Inc.
- Distributor
- Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden.
- Lot
- Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDACareLink Clinic, REF: MMT-7350October 28, 2026
- FDAGeneric Name: MiniMed Go App Generic Name: MiniMed Go App Model/Catalog Number: MMT-8600 (iOS), MMT-8601October 28, 2026
- FDABrand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072October 14, 2026
- FDAMiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;August 26, 2026
- FDAEmano Flow with software version v2.0.0 GS1 GTIN code: 199874903484 Emano Flow is a smartphone-based acousticJuly 29, 2026
- FDAMedline convenience kits containing Bard® Silastic® Foley Catheters (French sizes 16, 18, 20) 5cc balloons.July 29, 2026
Questions about this recall
Was MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android) recalled?
Yes. FDA published recall Z-2778-2026 on July 29, 2026. Reported hazard: For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.
What should I do if I have this product?
Voluntary: Firm initiated · Letter