Back to search
FDAFDA · Food and drugs·Recall number Z-2778-2026

MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android)

Published July 29, 2026 · Recall initiated: June 16, 2026

Summary

FDA records medical device Class II recall Z-2778-2026, published July 29, 2026 and initiated June 16, 2026. Recalling firm: Medtronic MiniMed, Inc.; Status: Ongoing; Reported quantity: 3,465; Lot: Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic MiniMed, Inc.
Distributor
Worldwide - US Nationwide distribution in the states and territories of PR, OH, WA, IL, NC, PA, VA, MN, ID, CA, UT, MI, NJ, MS, SC, CO, AZ, OR, AR, MO, TX, AL, FL, NY, IN, TN, IA, WI, MA, LA, OK, GA, KS, NM, CT, KY, NH, NV, MD and the countries of Australia, Switzerland, Germany, Denmark, Spain, Finland, Iceland, Norway, Sweden.
Lot
Model/UDI-DI: MMT-8600/00199150038978(10)1.1.0, MMT-8601/00199150038985(10)1.1.0 Software Versions: 1.0.0, 1.1.0
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was MiniMed Go App, Model: MMT-8600 (iOS), MMT-8601 (Android) recalled?

Yes. FDA published recall Z-2778-2026 on July 29, 2026. Reported hazard: For MiniMed Go app version 1.0.0 and 1.1.0, when the Set day and night is turned OFF and missed dose or Correct high glucose alerts are enabled, the associated alerts will not be triggered during the nighttime. This could result in missed rapid-acting insulin doses without the user s knowledge, and may lead to delayed therapy and an increased risk of hyperglycemia.

What should I do if I have this product?

Voluntary: Firm initiated · Letter