EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78222
Published August 5, 2026 · Recall initiated: June 24, 2026
Summary
FDA records medical device Class II recall Z-2810-2026, published August 5, 2026 and initiated June 24, 2026. Recalling firm: Medtronic Perfusion Systems; Status: Ongoing; Reported quantity: 3,621 units; Lot: UDI-DI: 20763000135700, Lot Numbers: 0232602300 0232602383 0232612916 0232680078 0232708967 0232709001 0232910880 0232910881 0232910882 0233077983 0233245850 0233245860 0233245980; UDI-DI: 00763000135706, Lot Numbers: 0232612916 0232709001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Perfusion Systems
- Distributor
- Worldwide - US Nationwide distribution.
- Lot
- UDI-DI: 20763000135700, Lot Numbers: 0232602300 0232602383 0232612916 0232680078 0232708967 0232709001 0232910880 0232910881 0232910882 0233077983 0233245850 0233245860 0233245980; UDI-DI: 00763000135706, Lot Numbers: 0232612916 0232709001
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAEOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as anAugust 5, 2026
- FDAEOPA Elongated One-Piece Arterial Cannula BLUNT 22FR, intended for use with cardiopulmonary bypass as anAugust 5, 2026
- FDAEOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as anAugust 5, 2026
- FDAEOPA Elongated One-Piece Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as anAugust 5, 2026
- FDAEOPA 3D Arterial Cannula 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula,August 5, 2026
- FDAAMS ALIGNED MEDICAL SOLUTIONS medical convenience kits labeled as: SPINAL KIT, Part Number AMS13683AAugust 5, 2026
Questions about this recall
Was EOPA 3D Arterial Cannula DIL TIP 22FR, intended for use with cardiopulmonary bypass as an arterial return cannula, Model Number 78222 recalled?
Yes. FDA published recall Z-2810-2026 on August 5, 2026. Reported hazard: Pinhole leaks may occur in the wire-wound body during cardiopulmonary bypass. Potential risks associated with this condition may include hypovolemia, vessel laceration or perforation, hypotension, ischemia, and organ dysfunction.
What should I do if I have this product?
Voluntary: Firm initiated · Letter