Back to search
FDAFDA · Food and drugs·Recall number Z-2823-2026

Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer system

Published August 5, 2026 · Recall initiated: May 8, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Instrumentation Laboratory
Distributor
US Nationwide distribution.
Lot
GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Hong Kong: GEM Premier 5000 Serial Nos.: 20106073, 20106039, 19052933, 20116415, 20064792, 22062065, 22062067, 19103490, 22092716, 22102881, 18112339, 19012518, 20126582, 20075256, 20064794, 20106036, 24016057, 21088758, 21088759, 18061910, 24016081, 24016108, 24108036, 24108044, 24067370, 24067371 and 24077468. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
Risk classification
Class II
Area (FDA)
device