Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer system
Published August 5, 2026 · Recall initiated: May 8, 2026
Summary
FDA records medical device Class II recall Z-2823-2026, published August 5, 2026 and initiated May 8, 2026. Recalling firm: Instrumentation Laboratory; Status: Ongoing; Reported quantity: 27; Lot: GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Hong Kong: GEM Premier 5000 Serial Nos.: 20106073, 20106039, 19052933, 20116415, 20064792, 22062065, 22062067, 19103490, 22092716, 22102881, 18112339, 19012518, 20126582, 20075256, 20064794, 20106036, 24016057, 21088758, 21088759, 18061910, 24016081, 24016108, 24108036, 24108044, 24067370, 24067371 and 24077468. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Instrumentation Laboratory
- Distributor
- US Nationwide distribution.
- Lot
- GEM Premier 5000: Part number: 00024019255 UDI: 08426950870227 Units distributed in Hong Kong: GEM Premier 5000 Serial Nos.: 20106073, 20106039, 19052933, 20116415, 20064792, 22062065, 22062067, 19103490, 22092716, 22102881, 18112339, 19012518, 20126582, 20075256, 20064794, 20106036, 24016057, 21088758, 21088759, 18061910, 24016081, 24016108, 24108036, 24108044, 24067370, 24067371 and 24077468. Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)October 21, 2026
- FDAHemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009800127. in vitro diagnosticOctober 21, 2026
- FDAGEM Premier 7000 with iQM3 Part Numbers: 00077407508 00077407510 00077415004 00077415008 00077415010August 12, 2026
- FDAGEM Premier 5000 PAK Part Numbers: 00055360010 00055360011 00055407504 00055415010 00055430010 00055430011August 12, 2026
- FDABrand Name: GEM Premier 7000 with iQM3 Product Name: GEM Premier 7000 with iQM3 Model/Catalog Number:August 5, 2026
- FDABrand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable bloodAugust 5, 2026
Questions about this recall
Was Brand Name: GEM Premier 5000 Product Name: GEM Premier 5000 Model/Catalog Number: 00024019255 portable blood analyzer system recalled?
Yes. FDA published recall Z-2823-2026 on August 5, 2026. Reported hazard: Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
What should I do if I have this product?
Voluntary: Firm initiated · Letter