Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i C20 6888520 Flow-i C30 6888530 Flow-i C40 6888540 Flow-c 6887700 Flow-e 6887900
Published August 12, 2026 · Recall initiated: June 24, 2026
Summary
FDA records medical device Class II recall Z-2860-2026, published August 12, 2026 and initiated June 24, 2026. Recalling firm: Maquet Critical Care AB; Status: Ongoing; Reported quantity: 408 US; 11,342 OUS; Lot: Flow-i C20 6677200 / UDI-DI 07325710001349 Flow-i C30 6677300 / UDI-DI 07325710001363 Flow-i C40 6677400 / UDI-DI 07325710001318 Flow-i C20 6888520 / UDI-DI 07325710010617 Flow-i C30 6888530 / UDI-DI 07325710010624 Flow-i C40 6888540 / UDI-DI 07325710010631 Flow-c 6887700 / UDI-DI 07325710009765 Flow-e 6887900 / UDI-DI 07325710010457 Software versions 4.8, 4.10, and 4.11. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Critical Care AB
- Distributor
- Worldwide - US Nationwide distribution to the states of AL, AR, CA, CO, DC, DE, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, SC, TX, VA, WA, WV.
- Lot
- Flow-i C20 6677200 / UDI-DI 07325710001349 Flow-i C30 6677300 / UDI-DI 07325710001363 Flow-i C40 6677400 / UDI-DI 07325710001318 Flow-i C20 6888520 / UDI-DI 07325710010617 Flow-i C30 6888530 / UDI-DI 07325710010624 Flow-i C40 6888540 / UDI-DI 07325710010631 Flow-c 6887700 / UDI-DI 07325710009765 Flow-e 6887900 / UDI-DI 07325710010457 Software versions 4.8, 4.10, and 4.11.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Sweden
Related recalls
- FDAGETINGE Y SENSOR Module, SERVO-U, SERVO-N, Model Number 6693662; part of the SERVO-U/-N continuous ventilatorAugust 26, 2026
- FDAAnti-Torque, Vector, Part: LS-749 part of Vector Pedicle Screw SystemAugust 12, 2026
- FDABrand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002August 12, 2026
- FDABrand Name: Medical Action Industries Product Name: Dialysis Change Tray Clear Sequence Model/Catalog Number:August 12, 2026
- FDASTRYKER AHTO TUBE SET PACKAGING, Catalog Number 0250070600August 12, 2026
- FDABrand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1August 12, 2026
Questions about this recall
Was Flow Anesthesia Machine Flow-i C20 6677200 Flow-i C30 6677300 Flow-i C40 6677400 Flow-i C20 6888520 Flow-i C30 6888530 Flow-i C40 6888540 Flow-c 6887700 Flow-e 6887900 recalled?
Yes. FDA published recall Z-2860-2026 on August 12, 2026. Reported hazard: Events observed where anesthetic agent was not delivered when the vaporizer was activated. Failure to deliver the anesthetic agent may result in the patient receiving insufficient anesthesia and may lead to intraoperative awareness.
What should I do if I have this product?
Voluntary: Firm initiated · Letter