ROSA Brain, 3.0.0 The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate.
Published August 9, 2017 · Recall initiated: October 11, 2016
Summary
FDA records medical device Class II recall Z-2868-2017, published August 9, 2017 and initiated October 11, 2016. Recalling firm: Zimmer Biomet, Inc.; Status: Terminated; Reported quantity: 20 units (16 US, 4 OUS); Lot: Device Listing No. D264591, Serial No. BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, BR16021, BR16023, BR16024. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Complaint of the head holder connector locking up mechanically when tightened.
What to do
Voluntary: Firm initiated · Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer Biomet, Inc.
- Distributor
- Worldwide Distribution - US including CA, FL, MA, MN, NC, NY, OH, PA, DC; Foreign - Australia, and France
- Lot
- Device Listing No. D264591, Serial No. BR16004, BR16005, BR16006, BR16009, BR16010, BR16011, BR16012, BR16013, BR16014, BR16015, BR16016, BR16017, BR16018, BR16021, BR16023, BR16024.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASpinal Rod Cutter, Item number 00392500200. The product is a nonsterile pin cutter - Product Usage: TheDecember 30, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Snap-In O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 32 mm, Neutral, O.D. Cup with Spacers, 55November 11, 2020
- FDAZCA All Poly Acetabular Cup Longevity Crosslinked Polyethylene, I.D. 26 mm, 10 degree inclined face, O.D. CupNovember 11, 2020
- FDAMULTIGENT Creatinine (Enzymatic), Catalog Numbers 8L24-31 and 8L24-41; IVD. Intended for the quantitativeAugust 9, 2017
- FDANon-Locking screws in the following lengths: (a) NEO(R) 0.27 - 10 mm, REF W27 ST010 (b) NEO(R) 0.27 - 12 mm,August 9, 2017
Questions about this recall
Was ROSA Brain, 3.0.0 The device is intended for the spatial positioning and orientation of instrument holders or tool guides to be used by neurosurgeons to guide standard neurosurgical instruments (biopsy needle, stimulation or recording electrode, endoscope). The device is indicated for any neurosurgical procedure in which the use of stereotactic surgery may be appropriate. recalled?
Yes. FDA published recall Z-2868-2017 on August 9, 2017. Reported hazard: Complaint of the head holder connector locking up mechanically when tightened.
What should I do if I have this product?
Voluntary: Firm initiated · Visit