Back to search
FDAFDA · Food and drugs·Recall number Z-2873-2026

Brand Name: Elevaris/Spectra Medical Devices Product Name: Sodium Chloride Injection, 0.9%, USP 10mL Ampule Model/Catalog Number: AMPUL10MLK. Sodium Chloride Flush for IV tubing systems and indwelling intravascular access devices.

Published August 19, 2026 · Recall initiated: July 6, 2026

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Manufacturing firm, Huons, was cited for issues regarding product quality and sterility assurance. If the product sterility is compromised, patients may be exposed to microorganisms, endotoxins, or foreign matter. This may trigger inflammatory responses, infections, micro-clots, and organ failure.

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Spectra Medical Devices, Llc
Distributor
Distributed in the United States to medical device kit packers in six (6) states - CA, FL, NC, OK, TX, VA.
Lot
Lot Code: Product codes: 1510-1 (individual ampule) / 1510-2 (carton) Lots: AB3005, AB3006, AB3007, AB3008, AB3009, AB3010, AB4001, AB4002, AB4003, AB4004, AB4005 AB4006, AB5001, AB5002, AB5003, AB5004, AB5005,AB5006, AB5007, AB5008 No individual UDI, as this product is used only within convenience kits.
Risk classification
Class I
Area (FDA)
device
Country
United States