Flexible Intubation Fiberscope, Part: 11302BD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube.
Published September 2, 2020 · Recall initiated: July 27, 2020
Summary
FDA records medical device Class II recall Z-2893-2020, published September 2, 2020 and initiated July 27, 2020. Recalling firm: Karl Storz Endoscopy; Status: Ongoing; Reported quantity: 650; Lot: Instruction Manual Version: 08/2018. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Karl Storz Endoscopy
- Distributor
- Worldwide distribution - U.S. Nationwide distribution including in the states of CA, NY, PA, AZ, OH, NJ, FL, NC, IL, SC, MA, IA, ND, WI, UT, AR, GA, WA, KY, CO, TX, TN, AL, LA, NE, WV, MO, MN, VA, MT, CT, MI, IN, NM, NV, ME, VT, NH, WY, DC, DE, OR, OK, MD, ID, AK, MS, KS, HI, RI, SD and the country of Canada.
- Lot
- Instruction Manual Version: 08/2018
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHOPKINS Telescope 30®, 4mm REF 10005BA UDI-DI code: 04048551045121 KARL STORZ HOPKINS® and HOPKINS® IIOctober 7, 2026
- FDAFlexible Hystero-Fiberscope Kits REF: 11264BBU1, 11264BBUK1, that contain expired Single Use Endoscopic Seal:July 29, 2026
- FDAKARL STORZ - ENDOSKOPE, REF: 27010L, Uretero-Renoscope, 7 Fr., 43cm, NON STERILE, RxONLY, CE 0123February 5, 2025
- FDAKARL STORZ - ENDOSKOPE, REF: 27003L, MICHEL Uretero-Renoscope, 9.5 Fr., 43cm, RxONLY, CE0123February 5, 2025
- FDAFlexible Intubation Fiberscope, Part: 11301BN1, with Instruction Manual: Z18441US - Product Usage: areSeptember 2, 2020
- FDAFlexible DCI Intubation Fiberscope, Part: 11302BDD2, with Instruction Manual: Z18445US - Product Usage: areSeptember 2, 2020
Questions about this recall
Was Flexible Intubation Fiberscope, Part: 11302BD2, with Instruction Manual: Z18445US - Product Usage: are intended to be used for examination and visualization of a patient s upper airway and for aiding placement of an endotracheal tube. recalled?
Yes. FDA published recall Z-2893-2020 on September 2, 2020. Reported hazard: Updated sterilization instructions are being provided to ensure 1) a specific sterilization tray is not used, and 2) ethylene oxide sterilization is never used on certain reusable flexible intubation endoscopes with total lumen lengths greater than 850mm. Incompletely sterilized endoscopes might introduce infectious agents into patient airways, resulting in infection.
What should I do if I have this product?
Voluntary: Firm initiated · Letter