FDAFDA · Food and drugs·Recall number Z-2909-2020
Alaris System PC Unit Model 8015
Published September 16, 2020 · Recall initiated: August 4, 2020
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The pump module keypad may exhibit keys that are unresponsive or stuck as a result of fluid ingress potentially resulting in a delay to the start of infusion, interruption of infusion, or inability to titrate medication. For high-risk patient populations who are receiving life-sustaining infusions, delays in an infusion or interruption can cause serious injury or death.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- CareFusion 303, Inc.
- Distributor
- U.S.:LA, FL, KY, MI, AL, GA, CA, TN, NC, TX, MA, OH, CO, WI, IL, KS, MO, UT, AK, MD, NY, NV, VA, NE, IN, PA, VT, DC, SC, CT, AR, NJ, MT, ND ,HI , NH, WA, DE, ME, OR, MN, AZ, MS, IA, NM, WY, ID, OK, SD, WV, MP, GU O.U.S.: CA, AE, AU, BE, CA, CH, IL, IN, KW, N/A, NL, NZ, PH, QA, SA, SG, TW, US, ZA
- Lot
- All serial numbers manufactured from April 07, 2017 to Present.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Sell products like this?