PROFEMUR Neck Long Titanium, Catalog Numbers: PHA01204, PHA01214, PHA01224, PHA01234, PHA01244, PHA01254, PHA01264. Hip prosthesis component.
Published September 30, 2020 · Recall initiated: July 31, 2020
Summary
FDA records medical device Class I recall Z-2941-2020, published September 30, 2020 and initiated July 31, 2020. Recalling firm: MicroPort Orthopedics Inc.; Status: Terminated; Reported quantity: 123,284 units; Lot: All lots/serial numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR® Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- MicroPort Orthopedics Inc.
- Distributor
- worldwide, except China and Japan
- Lot
- All lots/serial numbers
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAEVOLUTION MP FEMORAL COMPONENT CS/CR POROUS SIZE 5 RIGHT PRIMARY, REF: EFSRP5PR; EVOLUTION MP FEMORALDecember 25, 2024
- FDAEVOLUTION® MP Femoral CS/CR Non-Porous Non-porous MicroPort total knee replacement implants are for cementedJuly 10, 2024
- FDACadia Organic Unsalted Microwave Popcorn, Net Wt. 8.4 oz (238 g) 3-2.8 oz (79 g) bags, KeHDistributors LLC,September 30, 2020
- FDADxA 5000; Catalog Number:B50516; UDI:15099590707118 - Product Usage: The DxA 5000 is a high-speed, modular,September 30, 2020
- FDAPROFEMUR Titanium and Cobalt Chrome modular necks, Item Numbers and descriptions: Item Number DescriptionSeptember 16, 2020
- FDAMicroPort Orthopedics SCHANZ SCREW SELF DRILL, SIZE 4mm, REF 20070057, STERILE R Product Usage: MPO Total HipNovember 1, 2017
Questions about this recall
Was PROFEMUR Neck Long Titanium, Catalog Numbers: PHA01204, PHA01214, PHA01224, PHA01234, PHA01244, PHA01254, PHA01264. Hip prosthesis component. recalled?
Yes. FDA published recall Z-2941-2020 on September 30, 2020. Reported hazard: MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR® Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
What should I do if I have this product?
Voluntary: Firm initiated · Letter