Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number 115-066755-00
Published September 11, 2024 · Recall initiated: July 22, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Mindray DS USA, Inc. dba Mindray North America
- Distributor
- US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada
- Lot
- UDI-DI: 06936415975784 Serial Number: AH6D28000517, AH6D28000518
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States