FDAFDA · Food and drugs·Recall number Z-2949-2024

Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number 115-066755-00

Published September 11, 2024 · Recall initiated: July 22, 2024

Summary

FDA records medical device Class II recall Z-2949-2024, published September 11, 2024 and initiated July 22, 2024. Recalling firm: Mindray DS USA, Inc. dba Mindray North America; Status: Ongoing; Reported quantity: 2 units; Lot: UDI-DI: 06936415975784 Serial Number: AH6D28000517, AH6D28000518. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Mindray DS USA, Inc. dba Mindray North America
Distributor
US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada
Lot
UDI-DI: 06936415975784 Serial Number: AH6D28000517, AH6D28000518
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number 115-066755-00 recalled?

Yes. FDA published recall Z-2949-2024 on September 11, 2024. Reported hazard: V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

What should I do if I have this product?

Voluntary: Firm initiated · Letter