Back to search
FDAFDA · Food and drugs·Recall number Z-2950-2024

Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: 115-066756-00

Published September 11, 2024 · Recall initiated: July 22, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Mindray DS USA, Inc. dba Mindray North America
Distributor
US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada
Lot
UDI-DI: 06936415975777 Serial Number: AH6A28000471 AH6A24000202 AH6A28000474 AH6A27000378 AH6A28000466 AH6A24000195 AH6A28000457 AH6A25000237 AH6A28000426 AH6A24000199 AH6A24000207
Risk classification
Class II
Area (FDA)
device
Country
United States