FDAFDA · Food and drugs·Recall number Z-2953-2020

13 ft Handpiece Cable, component of the Dermatome Sets Model S and Model S6 and also as a spare part - Product Usage: are intended for use in the cutting of skin grafts for use in plastic surgery and/or burn treatment. These instruments are intended only for professional use in hospitals and similar medical facilities, where the patient will be under the supervision of trained personnel.

Published September 23, 2020 · Recall initiated: August 4, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

lntegra LifeSciences has identified through a number of complaints that the handpiece cable melts following sterilization.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Integra LifeSciences Corp.
Distributor
Worldwide distribution - US Nationwide distribution including in the states of MN, TX, CA, UT, WV, FL, IN, GA, AZ, and the countries of Latin America ( CR, MX, TT, EC, PA ), Canada, Europe (FR, AT, BE, ES).
Lot
Catalog Number: 353971513 Catalog Description: Handpiece Power Cable Model #: N/A Lot/Serial #: PT-3195
Risk classification
Class II
Area (FDA)
device
Country
United States