Impella CP Pump 371 Set, Clinical Trial; Product Code: 0048-0040;
Published September 11, 2024 · Recall initiated: August 5, 2024
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Abiomed, Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.
- Lot
- Product Code: 0048-0040; UDI-DI: 00813502011739; Serial Numbers: 370089 370091 370090 370074 370093 401804 401805 401811 401806; Batch Numbers: 2023172492 2023172497 2023172494 2023172500 2023172499 2023255511 2023255513 2023255522 2023255514
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States