FDAFDA · Food and drugs·Recall number Z-2967-2024

ImpellaCP SmartAssistSetAPAC, Delo1895; Product Code: 1000302;

Published September 11, 2024 · Recall initiated: August 5, 2024

Summary

FDA records medical device Class II recall Z-2967-2024, published September 11, 2024 and initiated August 5, 2024. Recalling firm: Abiomed, Inc.; Status: Ongoing; Reported quantity: 8 units; Lot: Product Code: 1000302; UDI-DI: 00813502012767; Serial Numbers: 392664 392661 392660 392665 392666 392667 392668 392669; Batch Numbers: 2023263121 2023263124 2023263125 2023269641 2023269670 2023269672 2023269673 2023269674. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Abiomed, Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei Darussalam, Canada, Croatia, Czeck Republic, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, Spain, Sweden, Switzerland, Taiwan, U.A.E., United Kingdom.
Lot
Product Code: 1000302; UDI-DI: 00813502012767; Serial Numbers: 392664 392661 392660 392665 392666 392667 392668 392669; Batch Numbers: 2023263121 2023263124 2023263125 2023269641 2023269670 2023269672 2023269673 2023269674;
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was ImpellaCP SmartAssistSetAPAC, Delo1895; Product Code: 1000302; recalled?

Yes. FDA published recall Z-2967-2024 on September 11, 2024. Reported hazard: A potential device interaction between the Shockwave Coronary IVL Catheter and left-sided Impella Pumps during coronary interventions. Currently there is no Abiomed labeling (IFU etc.) content related to this potential interaction, so Abiomed will update the IFU to include information on this potential device interaction.

What should I do if I have this product?

Voluntary: Firm initiated · Letter