FDAFDA · Food and drugs·Recall number Z-2986-2020
Perceval Single Use Accessory Kit, Size S, REF: ICV1349, Sterile EO, Rx Only, CE, UDI: 8022057015211 - Product Usage: he Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation.
Published September 30, 2020 · Recall initiated: July 29, 2020
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
LivaNova received complaints indicating the impossibility to collapse Perceval valve during procedure. The Dual Collapser is a component of the Perceval single use accessory kits used to reduce the Perceval Sutureless heart valve diameter before valve implant, allowing the Perceval implantation. The issue is relevant to specific lots of Dual Collapser size S/M.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Sorin Group Italia S.r.l.
- Distributor
- US Nationwide distribution including in the states of CA, FL, ID, ME, MN, MO, MS, NC, SD, OH, OR, PA, SD, TN, TX, VA, WA, WI and the countries of Canada, Japan, Germany, Netherland, Switzerland, Sweden, UK, Spain, Italy, France, Belgium, Finland, Greece, Slovenia, Malta, Poland, Romania, Croatia, Hungary, Serbia, Australia, Israel, Georgia, Iran, Kuwait, Oman, Quatar, Singapore, Thailand, Taiwan, Russia, South Africa.
- Lot
- Kit Lots: 2002200196, 2001300339, 2002130159
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Italy