HomeChoice Automated PD system Model numbers: 5C4471, 5C4471R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Published August 23, 2017 · Recall initiated: July 5, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Baxter Healthcare Corporation
Distributor
Worldwide Distribution
Lot
all codes
Risk classification
Class II
Area (FDA)
device
Country
United States