HomeChoice Automated PD system Model numbers: 5C4471, 5C4471R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.
Published August 23, 2017 · Recall initiated: July 5, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- Worldwide Distribution
- Lot
- all codes
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States