VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).
Published August 30, 2017 · Recall initiated: June 1, 2017
Summary
FDA records medical device Class II recall Z-3047-2017, published August 30, 2017 and initiated June 1, 2017. Recalling firm: Solta Medical Inc; Status: Terminated; Reported quantity: 4 units; Lot: Reference No. 110-0021; Shipping P/N P007021-03; Device P/N P006982-03; Serial No. VAS10-0134, VAS13-0440, VAS12-0203, 000TWY. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Solta Medical Inc
- Distributor
- US Distribution to the states of : AL, CA, CO, FL, NY, PA, & TX.
- Lot
- Reference No. 110-0021; Shipping P/N P007021-03; Device P/N P006982-03; Serial No. VAS10-0134, VAS13-0440, VAS12-0203, 000TWY.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAIQF Green Beans, Keep Frozen. 11 varieties, packaged under the following brands and sizes: 1) BountifulAugust 30, 2017
- FDAORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular lens surgeryAugust 30, 2017
- FDAWegmans Apple Crostata, Large, Net Wt. 23 oz. and 6 oz. (In-Store Bakery)August 30, 2017
- FDAORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgeryAugust 30, 2017
- FDAThe Comfy Cow Salted Caramel Ice Cream 1 pintAugust 30, 2017
- FDAMedtronic Curved Suction 90 9733451 EM ENT, reusable, Rx. The firm name on the label is Medtronic Navigation,August 30, 2017
Questions about this recall
Was VASER Amplifier, VASER 2.2 VASERlipo System, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy). recalled?
Yes. FDA published recall Z-3047-2017 on August 30, 2017. Reported hazard: Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit