Back to search
FDAFDA · Food and drugs·Recall number Z-3083-2024

Baxter EXACTAMIX Inlet, Vented, High Volume Inlet, REF H938174, for use in a pharmaceutical compounding device

Published September 25, 2024 · Recall initiated: August 20, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Baxter Healthcare Corporation has received increased customer reports of particulate matter in the Automated Compounding Device Inlets (disposable inlet), Product Codes H938173, H938174, H938175, H938176, used with the ExactaMix and ExactaMix Pro compounders. Particulate matter has been observed within the inlet primary packaging inlet components, including within the sterile fluid path tubing, before use.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Baxter Healthcare Corporation
Distributor
Worldwide distribution.
Lot
UDI/DI 00085412475790, Lot Numbers: 803799 and lower
Risk classification
Class I
Area (FDA)
device
Country
United States