Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only. Used with automated external defibrillators
Published September 20, 2017 · Recall initiated: August 8, 2017
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
What to do
Voluntary: Firm initiated · E-Mail
Recall details
- Manufacturer
- Leonhard Lang Medizintechnik GmbH
- Distributor
- Worldwide distribution. US distribution in the states of : FL & TX. Austria, China, France, Germany, Greece, Hungary, India, Italy, Panama, Philippines, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, United Arab Emirates, United Kingdom.
- Lot
- Model Number DF59N (lot numbers 41219-0778, 50403-0772, 41029-0770), DF59NC (lot numbers 50211-0974, 50402-0974, 40806-0972, 41203-0973, 50116-0975, 50527-0974). Expiration date 09/2017 to 05/31/2018.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- Austria