Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only. Used with automated external defibrillators
Published September 20, 2017 · Recall initiated: August 8, 2017
Summary
FDA records medical device Class I recall Z-3109-2017, published September 20, 2017 and initiated August 8, 2017. Recalling firm: Leonhard Lang Medizintechnik GmbH; Status: Terminated; Reported quantity: 8,249 units total; Lot: Model Number DF59N (lot numbers 41219-0778, 50403-0772, 41029-0770), DF59NC (lot numbers 50211-0974, 50402-0974, 40806-0972, 41203-0973, 50116-0975, 50527-0974). Expiration date 09/2017 to 05/31/2018. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Leonhard Lang Medizintechnik GmbH
- Distributor
- Worldwide distribution. US distribution in the states of : FL & TX. Austria, China, France, Germany, Greece, Hungary, India, Italy, Panama, Philippines, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, United Arab Emirates, United Kingdom.
- Lot
- Model Number DF59N (lot numbers 41219-0778, 50403-0772, 41029-0770), DF59NC (lot numbers 50211-0974, 50402-0974, 40806-0972, 41203-0973, 50116-0975, 50527-0974). Expiration date 09/2017 to 05/31/2018.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- Austria
Related recalls
- FDAWegmans Assorted Muffins 4-Pack-18 oz. (1 lb. 2 oz.) 510g packaged in clear plastic shell. UPC: 77890-22744 .September 20, 2017
- FDAToshiba Medical Radrex - i Digital Radiography X-ray system Radrex-i is intended for use in conjunction withSeptember 20, 2017
- FDAZlotoklos brand Ciastka maslane w czekoladzie (Butter cookies in chocolate), 12.3 oz (350 g). Product isSeptember 20, 2017
- FDASpectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK)September 20, 2017
- FDALeader Liquid Multivitamin Supplement for Infants and Toddlers , 1 2/3 oz (50 mL)September 20, 2017
- FDABD Needle 22GA 1-1/2in SafetyGlide(TM) Non Sterile, Bulk, REF 301701 The BD SafetyGlide(TM) needle is intendedSeptember 20, 2017
Questions about this recall
Was Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only. Used with automated external defibrillators recalled?
Yes. FDA published recall Z-3109-2017 on September 20, 2017. Reported hazard: Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail