Back to search
FDAFDA · Food and drugs·Recall number Z-3115-2017

Spectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consist of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide.

Published September 20, 2017 · Recall initiated: June 26, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The inside diameter of the 2.3mm small drill guide inserts have a smaller diameter than the specifications require.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Implant Direct Sybron Manufacturing, LLC
Distributor
Worldwide Distribution - USA (nationwide) Distribution to the states of : AZ, CA, CO, CT, FL, ID, IL, KS, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA., and to the countries of : Canada, Netherlands, Spain, Great Britain, Germany and Switzerland.
Lot
Lot 65249, Sequence Number HE G-HKRev01, Device Listing Number D245883.
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?