Back to search
FDAFDA · Food and drugs·Recall number Z-3118-2017

Sekisui Diagnostics Sekure Chemistry Acetaminophen L3K , Ref. 506-30. The firm names on the label are Sekisui Diagnostics (UK) Limited, United Kingdom, and Sekisui Diagnostics P.E.I. Inc., Canada. Used in laboratory testing.

Published September 20, 2017 · Recall initiated: July 20, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reagent carryover from Sekisui ACTH when utilized on the BEC AU 680 analyzer may lead to falsely elevated BEC Lipase recovery exceeding the total precision specification.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
Distribution US nationwide.
Lot
All lots
Risk classification
Class II
Area (FDA)
device
Country
United States