Evolution Upgrade 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: (1) 782117 (2) 782143
Published October 2, 2024 · Recall initiated: September 3, 2024
Summary
FDA records medical device Class II recall Z-3193-2024, published October 2, 2024 and initiated September 3, 2024. Recalling firm: Philips North America Llc; Status: Ongoing; Reported quantity: 3 units; Lot: 782117 (01)00884838099739(21) 782143 (01)00884838108660(21) Serial Numbers: 85675 85687 85695. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America Llc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Iceland, India, Indonesia, Iran, Iraq, Israel, Italy, Japan, Jordan, Kuwait, Latvia, Maldives, Mauritius, Mexico, Morocco, Netherlands, Pakistan, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Switzerland, Thailand, T®rkiye, United Kingdom, Utd.Arab.Emir., Uzbekistan, Vietnam.
- Lot
- 782117 (01)00884838099739(21) 782143 (01)00884838108660(21) Serial Numbers: 85675 85687 85695
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPhilips Medical Systems Technologies Ltd. Spectral CT 7500, Software V5.2.x, Product Code 728333October 7, 2026
- FDAAchieva 1.5T, Achieva 1.5T Conversion and Achieva 1.5T Initial system; Model Numbers (REF): 1. 781196 (AchievaSeptember 30, 2026
- FDAIngenia 3.0T and Ingenia 3.0T CX; Model Numbers (REF): 1. 781342(3.0T), 2. 781377(3.0T), 3. 782103(3.0T), 4.September 30, 2026
- FDAIngenia 1.5T, Ingenia 1.5T CX and Ingenia 1.5T S; Model Numbers (REF): 1. 781315(1.5T), 2. 781341(1.5T), 3.September 30, 2026
- FDABioPince Ultra Full Core Biopsy Instrument: REF: REF: 360-1080-01, 360-1080-02, 360-1080-03, 360-1580-01,October 2, 2024
- FDAIngenia 1.5T -For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/orOctober 2, 2024
Questions about this recall
Was Evolution Upgrade 3.0T - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal structure of the head, body or extremities, in any orientation. (REF) Numbers: (1) 782117 (2) 782143 recalled?
Yes. FDA published recall Z-3193-2024 on October 2, 2024. Reported hazard: Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.
What should I do if I have this product?
Voluntary: Firm initiated · Letter