Platinium VR DF4 1240, Model Number TDF035U, UDI/GTIN 08031527015460, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT).
Published October 4, 2017 · Recall initiated: July 24, 2017
Summary
FDA records medical device Class II recall Z-3231-2017, published October 4, 2017 and initiated July 24, 2017. Recalling firm: Sorin Group Italia SRL - CRF; Status: Terminated; Reported quantity: 72 units; Lot: All Serial Numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sorin Group Italia SRL - CRF
- Distributor
- US Nationwide Distribution
- UPC
- 08031527015460
- Lot
- All Serial Numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Italy
Related recalls
- FDAPlatinium DR DF4 1540, Model Number TDF036U, UDI/GTIN 08031527015477, Implantable cardioverter defibrillator,October 4, 2017
- FDAPlatinium CRT-D 1741, Model Number TDF037U, UDI/GTIN 08031527015484, Implantable cardioverter defibrillator,October 4, 2017
- FDAPercuvance Maryland Dissector Tool Tip The Percutaneous Surgical System with 5mm attachments is indicated forOctober 4, 2017
- FDAPercuvance Gripper Grasper Tool Tip The Percutaneous Surgical System with 5mm attachments is indicated for theOctober 4, 2017
- FDAOVATIO Models: VR 6250, DR 6550, CRT 6750. Product Usage: OVATIO VR 6250, OVATIO DR 6550, PARADYM VR 8250,May 24, 2017
- FDAINTENSIA Models: VR 124, DR 154, CRT-D 174 Product Usage: OVATIO VR 6250, OVATIO DR 6550, PARADYM VR 8250,May 24, 2017
Questions about this recall
Was Platinium VR DF4 1240, Model Number TDF035U, UDI/GTIN 08031527015460, Implantable cardioverter defibrillator, Biventricular or ventricular antitachycardia pacing, Dual or single chamber arrhythmia detection Product Usage: PLATINIUM VR is indicated for use in patients who are at high risk of sudden cardiac death due to ventricular tachyarrhythmias and who have experienced one of the following situations: - Survival of at least one episode of cardiac arrest (manifested by the loss of consciousness) due to ventricular tachyarrhythmia, - Recurrent, poorly tolerated sustained ventricular tachycardia (VT). recalled?
Yes. FDA published recall Z-3231-2017 on October 4, 2017. Reported hazard: There is a possibility of overconsumption of certain PLATINIUM Implantable Cardiac Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillators (CRT-Ds) following an ElectroStatic Discharge (ESD) during the implant surgery or a Magnetic Resonance Imaging (MRI) scan, and what actions to take.
What should I do if I have this product?
Voluntary: Firm initiated · Letter
How do I know if my unit is part of the recall?
Compare your product details with the notice. UPC: 08031527015460.