***This recall is being conducted due to an incomplete recall for RES 57829*** Curlin Infusion Administration Set, REF 340-4165, Sterile EO, The Curlin Ambulatory Infusion Pump System with disposable PVC Administration Tubing Sets and user accessories, provide a means for the volumetric delivery of fluids used in Parenteral, Enteral, Epidural, Subcutaneous and Intravenous applications.
Publicado el 24 de octubre de 2012 · Inicio del retiro: 26 de septiembre de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
This recall is the same issue for the FDA processed recall initiated by Moog Medical Devices on 1/25/11 which was related to customers reporting unclearable Air-In-Line alarm with specific lot numbers of Curlin Administration Sets used with Curlin Ambulatory Infusion Pumps. The lots associated with this recall were inadvertently omitted from the original recall which has since been terminated by
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- MOOG Medical Devices Group
- Distribuidor
- Nationwide Distribution including California
- Lote
- Product Code/Lot Number: 340-4166/CF1022906; 340-4128-V/CRF10082001; 340-4165/CF1022905.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States