Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0528-2025

smiths medical PORTEX Tracheal Tube Exchange Guide, 15Ch x 70cm, REF 14-504-68 and REF 14-504-68JP

Publicado el 4 de diciembre de 2024 · Inicio del retiro: 28 de octubre de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Smiths Medical has identified a potential for ingress of fluid into the device during reprocessing. The ingress takes place at the rear of the device between the end cap and the tube. This could lead to staining of the device or allowing the fluid to remain in the device. In addition, the Hypochlorite Solution (200ppm) and the 4% Acetic Acid disinfectants recommended in the IFU may be inadequate according to the disinfection standards for this type of device.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Smiths Medical ASD, Inc.
Distribuidor
Worldwide distribution.
Lote
REF 14-504-68, UDI/DI 35021239002586, Lot Numbers: 0003393, 0003394, 0003357, 0003358, 0003267, 0003368, 0003634, 0003254, 0003369, 0003165, 0003184, 0003196, 0003204, 0003206, 0003226, 0003256, 0003300, 0003307, 0003308, 0003314, 0003322, 0003359, 0003360, 0003361, 0003362, 0003363, 0003364, 0003377, 0003380, 0003635, 0003643; REF 14-504-68JP, UDI/DI 35021239002586, Lot Numbers: 0003369, 0003377, 0003393, 0003394, 0003634, 0003635, 0003643
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States