Flexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers and Guiding Sheaths are intended to introduce therapeutic or diagnostic devices into the vasculature, excluding coronary and neuro vasculature.

Publicado el 30 de diciembre de 2020 · Inicio del retiro: 24 de noviembre de 2020

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is an increased likelihood of the introducer sheath separating at the proximal bond site. If separation occurs during use, it could result in life-threatening adverse events. The potential adverse events that may occur include, but are not limited to increased procedural time, intervention to retrieve a separated segment, embolization occluding blood flow to a vital organ, vessel injury, and hemorrhage.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Cook Inc.
Distribuidor
US Nationwide Distribution including states of: AK, CT, IA, LA, MO, NH, OK, TN, WI, AL, DC, ID, MA, MS, NJ, OR, TX, WV, AR, DE, IL, MD, MT, NM, PA, UT, WY, AZ, FL, IN, ME, NC, NV, RI, VA, CA, GA, KS, MI, ND, NY, SC, VT, CO, HI, KY, MN, NE, OH, SD, WA and Global Distribution to: AU, SG, Austria, Belgium, Canary Islands, CN, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, Ireland, Italy, KR, Latvia, Lithuania, MY, Morocco, Netherlands, Norway, Poland, Portugal, Reunion, Russian Federation, Saudi Arabia, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, TW, TH, and United Arabs Emirates.
Lote
Lot Numbers: 13135874, 13141336, 13162743, 13168158, 13192089, 13214454, 13229690, 13237656, 13265339, 13297262, 13312708, 13320587, 13323491, 13342872, 13349041, 13363230, 13382710, 13407498, NS13140914, NS13219558, NS13221811, NS13246568, NS13265338, NS13291878, NS13291879, NS13312707, NS13342874, NS13349040, NS13360428, NS13372495
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States