PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Publicado el 16 de enero de 2013 · Inicio del retiro: 12 de diciembre de 2006
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Biomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Biomet, Inc.
- Distribuidor
- Worldwide Distribution -- US (nationwide) including the countries of , Canada, Japan, Argentina, Australia ,Netherlands, Europe and Chile.
- Lote
- Part 113950 Lot 093540, 093550,093570,093580,093590,093600,093610,093620,093720, 093730,093740,093760,093770,093780,282590,282600,282620,282630, 282680,282700,282710,489160,489180,489200,489210,489770,489780, 489790,714480,714550,714570,714700,714730,722210,722250,722280, 984770,984780,984790,984800,984810,984820,984830,984840, and 984850
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.