INFX-8000V Bi-Plane X-Ray Interventional System X-ray systems

Publicado el 17 de febrero de 2016 · Inicio del retiro: 14 de agosto de 2015

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

When a fontal DA (Digital Angiography) acquisition was done, scattered x-ray came into the dose meter on the lateral side. As a result, the dose meter sent a minus value to the software. Consequently, the software defined the data as an "abnormal value" and it stopped displaying dose data and the dose data was lost. The following message was displayed, "Dose meter abnormal, Dose info disabled".

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Toshiba American Medical Systems Inc
Distribuidor
US Nationwide Distribution to OH and NY
Lote
Serial Numbers ACA1492002, AAA1492001, W4B1442020, and W4B1432018.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States