Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-0938-2014

Malibu Polyaxial Screwdriver, 8.0 Plus, Model Number 91-2109; Malibu Polyaxial Screwdriver, Standard, Model Number 91-2110. 510(k): 91-2109 K061342. 91-2110 K051942. The Malibu" System is intended to be used as a temporary or permanent posterior, non-cervical implant to correct spinal disorders and provide stabilization of the spine to permit the biological process of spinal fusions to occur. The System is comprised of single use pedicle screws of varying reduction height, rods, rod connectors, etc. and an accompanying set of reusable surgical instruments. Some of the surgical instruments are specific to the system and others are general surgical instruments. The polyaxial screwdriver is specific to the polyaxial screws in the system.

Publicado el 19 de febrero de 2014 · Inicio del retiro: 22 de enero de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Integra LifeSciences has identified through an investigation of complaints that there may be the potential for the shaft of the Malibu Polyaxial Drivers to break during use. None of the complaints reported a patient injury. However, some complaints did indicate the surgery was slightly prolonged while the additional driver provided in the kit was retrieved and used.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Integra LifeSciences Corp.
Distribuidor
Worldwide Distribution - USA including IA, FL, LA, IL, CA, TX, AZ, NC, MD, NM, VA, and Internationally to South Africa.
Lote
Model Number(s): 91-2109 (8.0 Plus) 91-2110 (Standard). Lot AL549784F: 91-2109 Malibu Polyaxial Screwdriver 8.0 Plus. Lot AL552404F: 91-2110 Malibu Polyaxial Screwdriver Standard. Lot AL546768F: 91-2110 Malibu Polyaxial Screwdriver Standard. The devices are not distributed as sterile and do not have a expiration date.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta