Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1017-2013

Medtronic Archer Super Stiff Guidewire 0.035 inch (0.89mm), Model ARCSJ200W. Intended to facilitate catheter placement and exchange during procedures in the aorta.

Publicado el 3 de abril de 2013 · Inicio del retiro: 13 de marzo de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Medtronic has identified an issue with a specific lot of the Archer 0.035 inch (0.89mm) Super Stiff Guidewire, model ARCSJ200W, where the PTFE (Teflon) coating has the potential to separate from the wire during in-vivo use.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Medtronic Inc. Cardiac Rhythm Disease Management
Distribuidor
Worldwide Distribution-USA including the states of FL, OH, and MN and the countries of Belgium, France Germany, Netherlands, Turkey, and United Kingdom.
Lote
Lot number GFWA0387
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States