Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries Model number: HT70 and HT70 Plus. The HT70 and HT70 Plus ventilator systems are intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces.

Publicado el 8 de mayo de 2013 · Inicio del retiro: 5 de abril de 2013

Resumen

FDA registra el retiro Z-1171-2013 de dispositivos médicos de Clase I, publicado el 8 de mayo de 2013 e iniciado el 5 de abril de 2013. Empresa responsable: Newport Medical Instruments Inc; Estado: Finalizado; Cantidad reportada: 2528 ventilators and 1864 power pac batteries; Lote: Model number: HT70 and HT70 Plus Power Pac battery serial number range: 2096134110001 to 2292048120050, Rev. C through Rev. F. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Newport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries due to customer reports of Newport HT70 and HT70 Plus ventilators alarming and going to internal backup battery sooner than expected while the ventilator is being operated on Power Pac battery.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Newport Medical Instruments Inc
Distribuidor
Worldwide Distribution - USA including US (AL, AR, AZ, CA, CO, CT, DE, FL, GA, lA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, TN, TX, UT, VA, WA, and WI. Internationally to Argentina, Australia, Bangladesh, Belgium, BoliVia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Ecuador, Egypt, France, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Ireland, Italy, Japan, Kazakhstan, Kosova, Lebanon, Lithuania, Malaysia, Mexico, Nambia, Nepal, Norway, Oman, Paraguay, Peru, Philippines, Poland, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Viet Nam.
Lote
Model number: HT70 and HT70 Plus Power Pac battery serial number range: 2096134110001 to 2292048120050, Rev. C through Rev. F
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries Model number: HT70 and HT70 Plus. The HT70 and HT70 Plus ventilator systems are intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1171-2013 el 8 de mayo de 2013. Peligro reportado: Newport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries due to customer reports of Newport HT70 and HT70 Plus ventilators alarming and going to internal backup battery sooner than expected while the ventilator is being operated on Power Pac battery.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter