Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries Model number: HT70 and HT70 Plus. The HT70 and HT70 Plus ventilator systems are intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces.
Publicado el 8 de mayo de 2013 · Inicio del retiro: 5 de abril de 2013
Resumen
FDA registra el retiro Z-1171-2013 de dispositivos médicos de Clase I, publicado el 8 de mayo de 2013 e iniciado el 5 de abril de 2013. Empresa responsable: Newport Medical Instruments Inc; Estado: Finalizado; Cantidad reportada: 2528 ventilators and 1864 power pac batteries; Lote: Model number: HT70 and HT70 Plus Power Pac battery serial number range: 2096134110001 to 2292048120050, Rev. C through Rev. F. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Newport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries due to customer reports of Newport HT70 and HT70 Plus ventilators alarming and going to internal backup battery sooner than expected while the ventilator is being operated on Power Pac battery.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Newport Medical Instruments Inc
- Distribuidor
- Worldwide Distribution - USA including US (AL, AR, AZ, CA, CO, CT, DE, FL, GA, lA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NH, NJ, NV, NY, OH, OK, OR, PA, Puerto Rico, RI, SC, TN, TX, UT, VA, WA, and WI. Internationally to Argentina, Australia, Bangladesh, Belgium, BoliVia, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Ecuador, Egypt, France, Greece, Hong Kong, Hungary, India, Indonesia, Israel, Ireland, Italy, Japan, Kazakhstan, Kosova, Lebanon, Lithuania, Malaysia, Mexico, Nambia, Nepal, Norway, Oman, Paraguay, Peru, Philippines, Poland, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Sri Lanka, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, Venezuela, and Viet Nam.
- Lote
- Model number: HT70 and HT70 Plus Power Pac battery serial number range: 2096134110001 to 2292048120050, Rev. C through Rev. F
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAThe GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in8 de mayo de 2013
- FDAMaquet Getinge Group battery modules for use with Maquet SERVO-i ventilator systems (Part #6487180). Battery8 de mayo de 2013
- FDAThe GemStar Pump is a small and lightweight, single channel infusion device designed for use in the home, in8 de mayo de 2013
- FDASTERRAD CYCLESURE 24 Biological Indicator (BI) The STERRAD CYCLESURE 24 Biological Indicator (BI), is intended8 de mayo de 2013
- FDAGemStar Pump - Battery Product Usage: The GemStar Pump is a small and lightweight, single channel infusion8 de mayo de 2013
- FDANewport HT70 and HT70 Plus Ventilators, Model Number: HT70 and HT70 Plus. Product Usage: The HT70 ventilator12 de septiembre de 2012
Preguntas sobre este retiro
¿Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries Model number: HT70 and HT70 Plus. The HT70 and HT70 Plus ventilator systems are intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or non-invasive interfaces. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1171-2013 el 8 de mayo de 2013. Peligro reportado: Newport Medical Instruments is conducting a voluntary recall on certain Newport Medical Instruments HT70 and HT70 Plus ventilator Power Pac batteries due to customer reports of Newport HT70 and HT70 Plus ventilators alarming and going to internal backup battery sooner than expected while the ventilator is being operated on Power Pac battery.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter