ECMO Reinforced Dual Lumen Cannula, Sterile EO, Rx Only, REF: VV28F

Publicado el 12 de junio de 2019 · Inicio del retiro: 17 de enero de 2017

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may potentially cause an immediate disconnect of the clear tube or lead to a later fatigue failure of the adhesive bond, with subsequent clear tube disconnect from the hub.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
OriGen Biomedical, Inc.
Distribuidor
US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland, South Africa, Chile
Lote
Lots: N18487, N18487-1, N18561-1, R19169
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States