LVP software of the Ivenix Infusion System (IIS), LVP-SW-0004

Publicado el 20 de marzo de 2024 · Inicio del retiro: 16 de septiembre de 2022

Estado reportado al momento de publicación: Completed — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.

Qué hacer

Voluntary: Firm initiated · Other

Detalles del retiro

Fabricante
Fresenius Kabi USA, LLC
Distribuidor
US Nationwide distribution in the states of WI and NJ.
Lote
UDI-DI 00811505030122 Software versions 5.0.1 and earlier.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States