Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1289-2024

STERIS Boot Assembly, Traction Device, OT 1000, Part Number REF 10009192

Publicado el 20 de marzo de 2024 · Inicio del retiro: 14 de febrero de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Three lots of the Boot Assembly, Traction Device are being recalled due to due to a customer report of an incorrectly oriented locking pin which caused the boot to detach from the traction system. If an impacted device is used during surgery, the boot may detach from the traction system, potentially leading to patient harm and a delay in surgery.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Baxter Healthcare Corporation
Distribuidor
US Nationwide distribution in the state of Alabama.
Lote
UDI/DI 00615521000851, Lot Numbers: 23-140, 23-268, 23-314
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta