MINI STICK MAX 4F X 10 CM STD .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-750

Publicado el 27 de marzo de 2024 · Inicio del retiro: 30 de enero de 2024

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Non-conformance may prevent the guidewire from passing through the introducer hub during a surgical procedure. The inability of the guidewire to pass through the introducer is due to the presence of voids in the internal lumen of the sheath hub, potential risk of the non-conformance is a delay in procedure, wherein the user may need to exchange the sheath to complete the case

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Angiodynamics, Inc.
Distribuidor
Worldwide distribution - US Nationwide and the countries of AT, CA, CH, CZ, ES, HK, KW, NL, NO, NZ, ZA.
Lote
UDI: 15051684022934 UPN: H965457501 Lot Number: 5784035 5784036 5784136 5787837 5788554 5788555 5789978 5790448 5790449 5791589 5791590 5791591 5797460 5797461 5798838 5799613 5799623 5800399 5804390 5806116
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States